Description
The 200-275 calorimeter uses the thermocouple technology developed for the small sam-
ple calorimeter series 601. This technique enables more than a decade increase in sensi-
tivity when compares to similar sized calorimeters.
Minimum power sensitivity of 10mW can be measured using its unique thermopile de-
sign. Powers up to 15 watts can also be achieved. The calorimeter utilises the calorime-
ters Peltier air bath technique to control the local environment.
User friendly software is included that will operate under NT4 or Windows 2000 and is
fully network compatible. Both end point power and equilibrium fitting software routines
are included in the standard package. As are decay correction for both tritium and pluto-
nium.
Measurement data is archived and can be analysed off line at any time.
The standard design can be easily adapted for different sample types or sample packag-
ing.
Specification
Thermal power measurement accuracy ·
Two versions are available. Unit can
·
better than 0.5% over the operating
be mounted on a single trolley or the
range and better that 0.2% at 1.0 Watts
control console and thermal element
power.
can be separated up to 10 meters
Custom design parameters means vari- ·
Weight of trolley mounted version is
·
able measurement chambers can be
300kG
accommodated
Power consumption 1000 watts maxi-
·
Measurement range from 1mW to 15
mum 110/230VAC
·
Watts