Description
The 350 series calorimeters determines the tritium mass in grams (g) and activity in units
of Curies (Ci) or Becquerels (Bq) by measuring the total heat output resulting from the
radioactive decay of tritium. The design has application to a wide range of measurement
requirements. These include in-plant accountancy measurements, shipper - receiver dif-
ference measurements, the identification of empty transport containers and international
safeguards measurements. The instrument also has a role in replacing a proportion of
measurements performed using the PVTc technique as the calorimetry measurement
process is non-intrusive.
The instrument is designed to measure tritium with a mass or activity ranging from over
10 g or 100 kCi (3.7 x 1015 Bq) down to less than 10 mg or 100 Ci (3.7 TBq). Measure-
ment uncertainty ranges from better than 0.2 % at 50 kCi to between 5 % and 10 % be-
low 1 kCi. Measurement times range from less than 2 hours to 4 - 5 hours depending on
the temperature of the getter and the quantity of tritium being measured.
The calorimetry measurement method is independent of sample pressure, chemical com-
position and the presence of any other non-radioactive material, for example hydrogen,
deuterium and helium. The measurement procedure is automated and requires almost no
manual intervention, with the exception of loading and unloading the calorimeter. A
calorimetric measurement of the tritium decay heat provides a direct measure of the total
tritium activity or tritium mass based on precise electrical power measurement.
User friendly software is included that operates under NT4 or Windows 2000 and is fully
network compatible. Both end point power and equilibrium fitting software routines are
included in the standard package, as are decay correction for both tritium and plutonium.
Measurement data is archived and can be analysed off line at any time.
Variations to the standard design for different requirements and for special sample types
or sample packaging can be accommodated.
Specification
Thermal power measurement accuracy ·
Two versions are available. Unit can
·
better than 1.0% over the operating
be mounted on a single trolley or the
range and better that 0.2% at 1.0
control console and thermal element
Watts power.
can be separated by up to 5 meters
Custom design parameters means vari- ·
Weight of trolley mounted version is
·
able measurement chambers can be
300kg
accommodated
Power consumption 1000 watts maxi-
·
Measurement range from 5mW to 20
mum 110/230VAC
·
Watts. The minimum power sensitivity
5mW with 100% RSD